CamAPS Glooko Review Tool
SUMMARY:
- CGM REVIEW: % Time CGM active: Target > 80%, Time in range (TIR): TIR target >70%, time below range (TBR): TBR target <4%
- INSULIN DEVICE: Total daily dose, basal/bolus split, how many boluses and? Overrides
- SYSTEM DETAILS: % Auto mode on, target >90%, use of “Boost” and “Ease-off”, carb entries

HYPERGLYCAEMIA (HIGH LEVELS) or TIR <70%
- Check infusion sites i.e., 2nd or 3rd daily cannula changes, check for lumps.
- Bolus related: ICR too weak, missed boluses, confidence with carb counting.
- High sensor glucose levels overnight or in between meals: optimise personal glucose target (PGT)
- ? Over treatment of hypoglycaemia – remember half usual glucose treatment.
- Consider using BOOST for sick days, post take away meals, premenstrual days.
HYPOGLYCAEMIA (LOW LEVELS) or TBR >3.9 mmol/L
- Bolus related: ICR too strong, late boluses, consider increasing PGT during times of greater variability.
- Activity related: consider using EASE OFF or disconnect pump. *Remember to enter hypo treatment into “ADD MEAL”